Solid microelectrodes for in situ voltammetric measurements
Pizeta, I.; Billon, G.; Fischer, J.-C.; Wartel, M. (2003). Solid microelectrodes for in situ voltammetric measurements. Electroanalysis 15(17): 1389-1396. https://dx.doi.org/10.1002/elan.200302669
In: Electroanalysis. Wiley-VCH: Weinheim. ISSN 1040-0397; e-ISSN 1521-4109, more
Peer reviewed article  

Available in  Authors 

Keyword
    Marine/Coastal

Authors  Top 
  • Pizeta, I., more
  • Billon, G.
  • Fischer, J.-C., more
  • Wartel, M., more

Abstract
    Solid Au and Ag microelectrodes for both,model,and in situ voltammetric measurements, have been designed andtested. Amalgamation procedure and the parameters of measurement procedure are discussed. Differences betweengold and silver electrodes were studied. Silver electrodes showed more stable behavior and were used prevalently for calibration and in situ measurements.Calibration of a 30-µm diameter Ag electrode for Mn(II), Fe(II), I(-I) andS(-II) showed good linearity with reasonable confidence intervals, necessary for quantification of the signalsobtained in situ, with average sensitivity of 0.05, 0.01, 0.16 and 1.10 nA/(µmol L-1) and minimal detection limit of 3.5, 28, 0.3 and 0.05 µmol L-1, respectively. Additional checking of the electrode sensitivity with the reduction peaks of oxygen is proposed.