Skip to main content

IMIS

A new integrated search interface will become available in the next phase of marineinfo.org.
For the time being, please use IMIS to search available data

 

[ report an error in this record ]basket (0): add | show Print this page

Coprecipitation with aluminium hydroxide and x-ray fluorescence determination of trace metals in water
Eltayeb, M.A.H.; Van Grieken, R. (1992). Coprecipitation with aluminium hydroxide and x-ray fluorescence determination of trace metals in water. Anal. Chem. 268(1): 177-183. http://dx.doi.org/10.1016/0003-2670(92)85262-5
In: Analytical chemistry. American Chemical Society: Washington. ISSN 0003-2700; e-ISSN 1520-6882, more
Peer reviewed article  

Available in  Authors 

Keyword
    Marine/Coastal

Authors  Top 
  • Eltayeb, M.A.H.
  • Van Grieken, R., more

Abstract
    Coprecipitation of trace elements on aluminium hydroxide was combined with energy-dispersive x-ray fluorescence for the analysis of environmental water. The parameters for the preconcentration method were studied. The optimum procedure involves addition of 1 mg of the carrier, at pH 7.3, to a 200-ml water sample. The pH was adjusted by addition of ammonia solution. Quantitative recoveries were obtained for Ti(IV), Cr(III_, As(V), Pb(II) and Th(IV). The precision was 4–6% and the detection limits were in the range 0.2–0.8 µg l-1.

All data in the Integrated Marine Information System (IMIS) is subject to the VLIZ privacy policy Top | Authors